The London Perl and Raku Workshop takes place on 26th Oct 2024. If your company depends on Perl, please consider sponsoring and/or attending.

NAME

Lab::Moose::Instrument::Lakeshore372 - Lakeshore Model 372 Temperature Controller

VERSION

version 3.792

SYNOPSIS

 use Lab::Moose;

 # Constructor
 my $lakeshore = instrument(
     type => 'Lakeshore372',
     connection_type => 'Socket',
     connection_options => {host => '192.168.3.24'},
     
     input_channel => '5', # set default input channel for all method calls
 );


 my $temp_5 = $lakeshore->get_T(channel => 5); # Get temperature for channel 5.
 my $resistance_5 = TODO

METHODS

get_T

my $temp = $lakeshore->get_T(channel => $channel);

 C<$channel> needs to be one of 'A', 1, ..., 16.

get_value

alias for get_T.

get_sensor_units_reading

 my $reading = $lakeshore->get_sensor_units_reading(channel => $channel);

Get sensor units reading (in ohm) of an input channel.

set_setpoint/get_setpoint

Set/get setpoint for loop 0 in whatever units the setpoint is using

 $lakeshore->set_setpoint(value => 10, loop => 0); 
 my $setpoint1 = $lakeshore->get_setpoint(loop => 0);

set_T

alias for set_setpoint

set_heater_range/get_heater_range

 $lakeshore->set_heater_range(loop => 0, value => 1);
 my $range = $lakeshore->get_heater_range(loop => 0);

For loop 0 (sample heater), value is one of 0 (off), 1, ..., 8. For loops 1 and 2, value is one of 0 and 1.

set_outmode/get_outmode

 $lakeshore->set_outmode(
  loop => 0, # 0, 1, 2
  mode => 3, # 0, ..., 6
  channel => 5, # A, 1, ..., 16
  powerup_enable => 0, # 0, 1
  polarity => 1, # 0, 1
  filter => 0, # 0, 1
  delay => 0, # 1,...,255
 );
 
 my $args = $lakeshore->get_outmode(loop => 0);

set_input_curve/get_input_curve

 # Set channel 5 to use curve 25
 $lakeshore->set_input_curve(channel => 5, value => 25);
 my $curve = $lakeshore->get_input_curve(channel => 5);

set_remote_mode/get_remote_mode

 $lakeshore->set_remote_mode(value => 0);
 my $mode = $lakeshore->get_remote_mode();

Valid entries: 0 = local, 1 = remote, 2 = remote with local lockout.

set_pid/get_pid

 $lakeshore->set_pid(loop => 1, P => 1, I => 50, D => 50)
 my %PID = $lakeshore->get_pid(loop => 1);
 # %PID = (P => $P, I => $I, D => $D);

set_zone/get_zone

 $lakeshore->set_zone(
     loop => 1,
     zone => 1,
     top  => 10,
     P    => 25,
     I    => 10,
     D    => 20,
     mout => 0, # 0%
     range => 1,
     rate => 1.2, # 1.2 K / min
     relay_1 => 0,
     relay_2 => 0,
 );

 my %zone = $lakeshore->get_zone(loop => 1, zone => 1);

set_filter/get_filter

 $lakeshore->set_filter(
     channel => 5,
     on      => 1,
     settle_time => 1, # (1s..200s) 
     window => 2, # % 2 percent of full scale window (1% ... 80%)
 );

 my %filter = $lakeshore->get_filter(channel => 5);

set_freq/get_freq

 # Set input channel 0 (measurement input) excitation frequency to 9.8Hz
 $lakeshore->set_freq(channel => 0, value => 1);

 my $freq = $lakeshore->get_freq(channel => 0);

Allowed channels: 0 (measurement input), 'A' (control input). Allowed values: 1 = 9.8 Hz, 2 = 13.7 Hz, 3 = 16.2 Hz, 4 = 11.6 Hz, 5 = 18.2 Hz.

Consumed Roles

This driver consumes the following roles:

Lab::Moose::Instrument::Common

COPYRIGHT AND LICENSE

This software is copyright (c) 2021 by the Lab::Measurement team; in detail:

  Copyright 2018       Simon Reinhardt
            2020       Andreas K. Huettel, Simon Reinhardt
            2021       Simon Reinhardt

This is free software; you can redistribute it and/or modify it under the same terms as the Perl 5 programming language system itself.