NAME
Lab::Instrument::HP34401A - HP/Agilent 34401A digital multimeter
SYNOPSIS
use Lab::Instrument::HP34401A;
my $hp=new Lab::Instrument::HP34401A(0,22);
print $hp->read_voltage_dc(10,0.00001);
DESCRIPTION
The Lab::Instrument::HP34401A class implements an interface to the 34401A digital multimeter by Agilent (formerly HP).
CONSTRUCTOR
my $hp=new(\%options);
METHODS
read_voltage_dc
$datum=$hp->read_voltage_dc($range,$resolution);
Preset and make a dc voltage measurement with the specified range and resolution.
- $range
-
Range is given in terms of volts and can be
[0.1|1|10|100|1000|MIN|MAX|DEF]
.DEF
is default. - $resolution
-
Resolution is given in terms of
$range
or[MIN|MAX|DEF]
.$resolution=0.0001
means 4 1/2 digits for example. The best resolution is 100nV:$range=0.1
;$resolution=0.000001
.
read_voltage_ac
$datum=$hp->read_voltage_ac($range,$resolution);
Preset and make an ac voltage measurement with the specified range and resolution. For ac measurements, resolution is actually fixed at 6½ digits. The resolution parameter only affects the front-panel display.
read_current_dc
$datum=$hp->read_current_dc($range,$resolution);
Preset and make a dc current measurement with the specified range and resolution.
read_current_ac
$datum=$hp->read_current_ac($range,$resolution);
Preset and make an ac current measurement with the specified range and resolution. For ac measurements, resolution is actually fixed at 6½ digits. The resolution parameter only affects the front-panel display.
display_on
$hp->display_on();
Turn the front-panel display on.
display_off
$hp->display_off();
Turn the front-panel display off.
display_text
$hp->display_text($text);
print $hp->display_text();
Display a message on the front panel. The multimeter will display up to 12 characters in a message; any additional characters are truncated. Without parameter the displayed message is returned.
display_clear
$hp->display_clear();
Clear the message displayed on the front panel.
beep
$hp->beep();
Issue a single beep immediately.
get_error
($err_num,$err_msg)=$hp->get_error();
Query the multimeter's error queue. Up to 20 errors can be stored in the queue. Errors are retrieved in first-in-first out (FIFO) order.
reset
$hp->reset();
Reset the multimeter to its power-on configuration.
id
$id=$hp->id();
Returns the instruments ID string.
CAVEATS/BUGS
probably many
SEE ALSO
AUTHOR/COPYRIGHT
This is $Id: HP34401A.pm 339 2006-04-12 17:30:48Z schroeer $
Copyright 2004-2006 Daniel Schröer (http://www.danielschroeer.de)
This library is free software; you can redistribute it and/or modify it under the same terms as Perl itself.
1 POD Error
The following errors were encountered while parsing the POD:
- Around line 182:
Non-ASCII character seen before =encoding in '6½'. Assuming CP1252